EXFO MAX700B - Retanking av OTDR

Kan gjøres i de tilfeller hvor OTDR nekter å starte eller om det er korrupte data.

  1. Perform a reset by holding the power button for > 12 seconds, the instrument should power up normally on the next power on
  2. If the instrument still does not boot correctly , try a second reset
  3. If the issue is not resolved, open the unit, and temporarily disconnect the battery. On next power on sequence, the instrument should boot normally
  4. If the problem persists, or if it happens frequently on any particular instrument, please report to EXFO support team

Doing this no data will be lost whereas a factory recovery will wipe all data. I would avoid this if possible.


Opening the unit to disconnect the battery will not void the product warranty.